-
Notifications
You must be signed in to change notification settings - Fork 1
New issue
Have a question about this project? Sign up for a free GitHub account to open an issue and contact its maintainers and the community.
By clicking “Sign up for GitHub”, you agree to our terms of service and privacy statement. We’ll occasionally send you account related emails.
Already on GitHub? Sign in to your account
No proper ppm scaling at different NMR-Machines #3
Comments
Hi Michael, Sorry for may answer delay. NMRProcFlow has been developed especially for Metabolomics approaches. So, we assumes that all spectra are acquired in the same conditions (instrument, sequence, solvent, buffer, pH, ppm range, ). Nevertheless, we also could compare different conditions in the same spectra set. Indeed, in this latter case, some issues can occurs. Is it possible to send me a very small subset of your spectra so that I can test and in this way could improve the software ? Regards |
Hi Daniel, |
Hi Michael, By using the 1r spectra, it assumes that the PPM calibration was made before within TopSpin. As it is not the case, we could add as the first macro-command : calibration -0.024 0.074 0 10.2 10.5 This would do the job. In addition I saw that you have two spectra sets very different (to say the least). I needed to normalize (CSN) the whole set within NMRProcFlow to see the second spectra set. There are a factor of 10^5 between them! I suppose you want to compare these two wine sets. Best wishes |
Hi Daniel, I have one comment though, when we look at the 1r spectra in TopSpin, the TSP peak is correctly set to zero for all spectra. And yes, you are right, comparing these spectra is a big challenge that we are still working on at the moment. Many greetings |
Hi Daniel,
we ran into a problem when reading in data from two different NMR spectrometers. Both are Bruker instruments, one of them with a NEO console. The problem is that the TSP signal is not zeroed correctly. Depending on which spectrometer is first in the listing, the spectra of the second spectrometer are shifted to it. The shift is about 0.04 ppm.
It seems that the TSP signal is not detected correctly. In the macro, only uniform bucketing is performed and no further processing steps are made.
Are we making a mistake in the reading?
With best regards
Michael
The text was updated successfully, but these errors were encountered: