According to the Bruker Biotyper database creation procedures mass spectra of the same technical replicate have to adhere to the following specifications:
mass range
max. allowed difference of masses [Da]
max. allowed deviation (500 ppm)
3000-4000
1.47
339.79
4000-5000
2.17
230.73
5000-6000
2.74
182.55
6000-7000
3.13
319.28
7000-8000
3.58
418.59
8000-9000
4.05
247.16
9000-10000
4.69
426.80
10000-11000
5.21
286.66
Measurement is done from the highest part of the peak in flexAnalysis. Doing this manually with Microsoft Excel is tedious and screams for automation or at least a nice visualization, which highlights differences of masses in a table or even better the selected peak with color.