This is a single-axis Galil-controlled device that has 3 pre-defined stop positions.
To set up the PGC, complete the following steps:
Go to C:\Instrument\Settings\Config\NDX[DEVICE_NAME]\configurations\galil
and create a file called motionsetpoints.cmd
if one doesn't already exist. Add the following:
$(IFDMC01) epicsEnvSet "LOOKUPFILE2" "$(ICPCONFIGROOT)/motionSetPoints/pgc.txt"
$(IFDMC01) motionSetPointsConfigure("LOOKUPFILE2","LOOKUPFILE2")
$(IFDMC01) dbLoadRecords("$(MOTIONSETPOINTS)/db/motionSetPoints.db","P=$(MYPVPREFIX)LKUP:PGC:,TARGET_PV1=$(MYPVPREFIX)MOT:MTR0102,TARGET_RBV1=$(MYPVPREFIX)MOT:MTR0102.RBV,TARGET_DONE=$(MYPVPREFIX)MOT:MTR0102.DMOV,TOL=1,LOOKUP=LOOKUPFILE2")
Don't forget to add a new line at the end of the final or the final line won't be read
Note that there are several points of customisation here:
1. The 01
in IFDMC01
refers to the Galil number. If you are using IOC GALIL_02
then this should instead be IFDMC02
1. References to MTR0102
should point to the motor you wish to use. If, for instance, you are using GALIL_06
, axis 3, this should be MTR0603
. Be careful to replace all the references, there are 3.
1. Make sure that the lookup file, LOOKUPFILE2
is unique in this file. You might have to use a difference name if LOOKUPFILE2
is already taken. So long as the name is consistent, it will work.
Create a file called C:\Instrument\Settings\Config\NDX[DEVICE_NAME]\configurations\motionSetPoints\pgc.txt
. You may have to create the directory as well if it doesn't exist already. Add the following:
Polariser -2
Guide 0
Collimator 2
There are several points of customisation:
- The names provide a guide to which position corresponds to which component. You can name these however you like and will be used to populate the buttons in the OPI.
- The second column is for the motor positions corresponding to each component. These will have to be set corresponding to the device setup.
Open a PGC OPI as normal either via a synoptic of the devices screen. Typical value for the macros are
-
PGC
:PGC
. This will only be different if you've changed the prefix inmotionsetpoints.cmd
. That is not recommended. -
MM
:MOT:MTR0102
. This corresponds to the motor record being used for the PGC and, combined with theMOT
prefix, will correspond to the value you used when setting upmotionsetpoints.cmd